Manufacturer's indications and declaration - electromagnetic emissions
The Timpani is designed for use in an electromagnetic environment as specified
below. The user of the Timpani must ensure that the instrument is operated in
ambient conditions as close as possible to these.
Emission tests
Conformity
Electromagnetic environment -
guidelines
RF emissions
CISPR11
Group 1
The Timpani utilizes RF energy for its
internal functions, and contains a Bluetooth
radio module that responds to pertinent
regulations. Consequently, the RF
emissions generated are minimal and
unlikely to interfere with other equipment
operating nearby.
RF emissions
CISPR11
Class B
The Timpani is suitable for use in
professional healthcare facilities and for
connection direct to the low voltage public
electricity grid.
Harmonic emissions
IEC 61000-3-2
Class A
Fluctuations in
voltage/emissions (flicker)
IEC 61000-3-3
Compliant
Manufacturer's indications and declaration - electromagnetic immunity
The Timpani is designed for use in an electromagnetic environment as specified
below. The user of the Timpani must ensure that the instrument is operated in
ambient conditions as close as possible to these.
Immunity tests
Test level to
IEC 60601
Level of
Compliance
Electromagnetic
environment -
guidelines
Electrostatic
discharges (ESD)
IEC 61000-4-2
± 8 kV contact
± 2 kV, ± 4 kV,
± 8 kV, ± 15 kV
air
± 8 kV contact
± 2 kV, ± 4 kV,
± 8 kV, ± 15 kV
air
The floor must be wood,
concrete or ceramic tile. If
floors are covered with
materials that present
synthetic coatings, relative
humidity must be at least
30%.
Electrical fast
transients / bursts
IEC 61000-4-4
± 2 kV for
electrical power
supply lines
± 1 kV for
input/output lines
± 2 kV for
electrical power
supply lines
± 1 kV for
input/output lines
The quality of electrical
power supplied by the grid
must correspond to that of
professional healthcare
facilities.
Overcurrent
IEC 61000-4-5
± 1 kV differential
mode
± 2 kV common
mode
± 1 kV differential
mode
± 2 kV common
mode
The quality of electrical
power supplied by the grid
must correspond to that of
professional healthcare
facilities.
Voltage dips, short
interruptions and
fluctuations
< 5% U
T
(1)
(dip >95% in U
T
)
< 5% U
T
(1)
(dip >95% in U
T
)
The quality of electrical
power supplied by the grid
must correspond to that of