3. Inspect the sensor to confirm that both wings of the electrode stack have engaged with the MiniPID
(4) body.
4. Refit the sensor into the sensing instrumentation.
5. The Falco must be calibrated
Use of PID Lamp Cleaning Kit A-31063
The container of cleaning compound contains Aluminium Oxide as a very fine power (CAS Number 1344-28-1).
The Compound has a TVL (TWA) of 10 mg/m
3
and a full material safety data sheet MSDS is available on request
from Ion Science Ltd. The key issues are listed below:
Hazard identification:
May cause irritation of respiratory tract and eyes.
Handling:
• Do not breathe the vapour/dust
• Avoid contact with skin, eyes and clothing
• Wear suitable protective clothing
• Follow industrial hygiene practices;
Wash face and hands thoroughly with soap and water after use and before eating, drinking, smoking or
applying cosmetics
• Always replace the lid after using the cleaning compound.
Storage:
Keep container closed to prevent water adsorption and contamination.
Fault Diagnostics
Alarm and Fault Indications
Activated when Alarm 1 is exceeded.
Activated when Alarm 2 is exceeded.
Activated when a fault condition is raised.