• State HIGH
• Expiration of the wait time [ms] Time
The parameters can be assigned in the following ways:
• Manual entry in the program editor
• Via the fieldbus interface (FBUS) for the event type Time.
GOTO
Goto command
Jumps to the defined command line and executes all subsequent com-
mands up to GOTO in a loop.
Parameter 1 indicates the number of the command line (0 – 99) as of
which the command set should be executed again.
Parameter 2 determines the event:
• SI HIGH: The command set defined in parameter 1 will be exe-
cuted as long as the signal HIGH is present at the synchronization
input.
• SI LOW: The command set defined in parameter 1 will be exe-
cuted as long as the signal LOW is present at the synchronization
input.
• Counter (0 – 32767): Number of repetitions of the command set
to be executed. With a value of 0, the defined command set is exe-
cuted in an endless loop.
OUT
Set signal “Synchronization Output”
• Sets signal LOW or HIGH at the output Sync. Out
• Parameter 1 selects a synchronization signal
• Parameter 2 sets the level (HIGH / LOW)
EXTPWR
Switch to external mode
This command specifies the external interface from which the value for
emission power and the time should be read for the commands SPT,
SPR and WAIT.
The execution of this command can be stopped with the commands
STOP, SPT, SPR and EXTPWR (by selecting a different external inter-
face).
SETCHAN
Selection of the beam switch channel
This command selects the optical channel for production with inte-
grated beam switch. The command can only be executed in the first line
of the program.
SETCHAN is only available for lasers with a beam switch.