Send the log files and the events file to support.europe@ipgphoton-
ics.com.
Message
Laser temperature too low
Type
Warning
Resetting
Automatic
Description
The warning Laser temperature too low indicates that the average mod-
ule temperature of all modules is below the permitted range limit of
19°C.
Possible      
causes
The laser stood for a prolonged period of time in a cold environ-
ment while the chiller was switched off.
The parameterization of the laser chiller is faulty.
The tank heater of the laser chiller is defective or not switched
on.
The laser chiller is defective (e.g. expansion valve).
Fault diagnos-
tics / trou-
bleshooting
ð Switch on the laser chiller. The heating elements of the laser
chiller heat the water to the necessary temperatures.
ð Check the parameterization of the laser chiller.
ð Check the tank heater of the laser chiller.
IPG chiller:
ð Using the Chiller tab, determine what type of fault this is.
ð See Messages of the Chiller tab or read the operating manual of
the laser chiller for a detailed fault analysis.
Message
Dew point temperature too high
Type
Warning
Description
ð See Messages of the Dew Point tab .
Message
Module(-s) without voltage supply
Type
Warning
Description
The warning Module(-s) without voltage supply is activated when no
voltage is present in the power circuit.
Possible      
causes
The laser power supply is not switched on.
The safety fuse of a module was tripped.
Analysis / trou-
bleshooting
ð Check whether the laser power supply is on (Laser ON).
ð Check the Modules tab to check the status of the modules (red or
gray background).
ð Check the safety fuses of all modules marked in red.