SPECIFICATIONS
49
SPECIFICATIONS
The specifications apply when the IPS X303 series are powered on for at least 30 minutes
under +20 to +30°C.
Output Ratings
CH1/CH2
Independent
0 to 30V / 0 to 3A
CH1/CH2 Series
0 to 60V / 0 to 3A
CH1/CH2 Parallel
0 to 30V / 0 to 6A
CH3
2.5V/3.3V/5.0V, 3A
CH4
0~5V,0~1A
Voltage Regulation
Line
0.01% + 3mV
Load
≤ 0.01% + 3mV (rating current ≤ 3A)
≤ 0.02% + 5mV (rating current > 3A)
Ripple & Noise
≤ 1mVrms (5Hz to 1MHz)
≤ 10mVpp (5Hz~1MHz)
Recovery Time
100µs (50% load change, minimum load 0.5A)
Temperature
Coefficient
300ppm/°C
Current Regulation
Line
0.2% + 3mA
Load
0.2% + 3mA
Ripple & Noise
3mArms
Tracking Operation
Tracking Error
≤ 0.1% + 10mV of Master (0~30V) (No Load, with
load add load regulation≤100mV)) (IPS X303S)
≤ 0.5%+50mV of Master (IPS 3303D)
Parallel
Regulation
Line: 0.01% + 3mV
Load: 0.01% + 3mV (rating current ≤ 3A)
Load: 0.02% + 5mV (rating current > 3A)
Series Regulation
Line: ≤ 0.01% + 5mV
Load: 300mV