ISOMAG ML 110 - DIGITAL INPUT; External Power Supply Digital Input; Internal Power Supply Digital Input; Input On;Off Operation

ISOMAG ML 110
36 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
11
0-1,5 V
3-40 V
AUTOCALIB. OFF
T
RESET
0-1,5 V
3-40 V
Tmin = 100ms
BLOCK
0-1,5 V
3-40 V
T
0-1,5 V
3-40 V
DIGITAL INPUT
INPUT ON/OFF OPERATION
3/40 Vdc (ON)
0/1.5 Vdc (OFF)
5 (+)
6 (-)
10 K
5
6
20
0
15
+24
10 K
Auto-calibration
Reset totalizers
Block totalizers
Block totalizers
Totalizers active
Necessary conditions for enabling the function
Necessary conditions for enabling the function
Necessary conditions for enabling the function
Tmin<T<1sec. = autocalibration
T > 1sec. = Auto zero
POS. 5.6 ENABLED (Autozero calibration external
command)
POS. 5.1 to 5.4 ENABLED (partial positive or reverse flow
totalizers reset enable).
POS. 5.5 ENABLED (totalizers counting lock command)
POS. 5.7 ENABLED (range change)
Necessary conditions for enabling the function
Scale 2
Scale 1
10Hz
20Hz
50Hz
80Hz
150Hz
220ms
110ms
45ms
30ms
15ms
T must be > Tmin
SAMPLE RATE
Tmin
SAMPLE RATE
Tmin
External power supply
Internal power supply