Power Supply Function
When the instrument is in Sense Reverse Protection state, you should check
whether the polarities are connected reversely or not firstly. If yes, you can re-
open the output after the polarities connect correctly.
The voltage difference between output terminal and remote sense terminal of
each model is not the same. When the remote sense terminal is connected re-
versely, the maximum voltage will not exceed the sum of output terminal voltage
and the difference voltage.
When Sense is reversed or short-circuited, the voltage meter value is dis-
played as a positive/negative value close to 0, and abnormal high voltage
output does not occur, which can avoid damage to the DUT.
4.6 Function Menu for Power Supply
The Function menu of the power supply includes the following:
Function Function Menu for Power Supply
LIST LIST program editing, file import/export, pro-
gram running (see 4.6.1 LIST Function).
BATTERY Battery charging testing function (see 4.6.2
Battery Charging Test Function).
4.6.1 LIST Function
The IT6000D series power supply supports a total of 10 List files (List01 to
List10), each of which can be set up to 200 steps. You need to edit the voltage/
current value, slope and time width of each step, or you can set repeat times (1
to 65535) for each List file. After the file is edited completely, you can trigger the
selected List file to run according to the selected trigger mode.
The descriptions of list function menu are listed in the table below.
LIST List menu
Run Indicates to enter the List run mode, waiting for the trigger to
run the currently selected List file.
Open Select the List file to open.