Calibration Command 
  Copyright©  Itech Electronics Co., Ltd.  98
Example 
CAL:VOLT 12.0V 
CALibrate:CURRent:LEVel 
This command is used to define the current calibration points. 
Command syntax 
CALibrate:CURRent:LEVel <P0|P1> 
Arguments 
P0|P1 
Example 
CAL:CURR:LEV P0 
CALibrate:CURRent[:DATA] 
This  command  is  used  to  read  the  actual  current  of  the  present 
calibration point. 
Command syntax 
CALibrate:CURRent[:DATA] <NRf> 
Arguments 
NRf 
Unit 
A 
Example 
CAL:CURR    20.0 
CALibrate:MONitor:VOLTage:LEVel 
This command is used to set monitoring the voltage calibration point. 
Command syntax 
CALibrate:MONitor:VOLTage:LEVel <P0|P1> 
Arguments 
P0|P1 
Example 
CAL:MON:VOLT:LEV P0 
CALibrate:MONitor:VOLTage[:DATA] 
This command is used to set the voltage calibration point data for monitoring.