System Commands 
  Copyright ©ITECH Electronic Co., Ltd.  17 
SYSTem:ERRor? 
This order is used to query the error information of the power supply. When the 
error LED lights on the front panel, it tells us that there is one or more hardware 
or order syntax error happened. There are at most 20 groups of error message. 
Send the order once will read one error message from the error queue. 
  Error messages follow the FIFO (first-in-first-out) principle. The first error 
will be returned first. when you read all the error tips information in the error 
queue. The ERROR led will turn off.    When error occur the buzzer of the 
power supply will beep once. 
  If more than 20 errors occurred, the last one stored in the queue will be 
replaced by “-350”, 
Meaning "too many mistakes.". If the error 
messages in the error message queue are not read out, the error 
following will not be saved to the queue. If there is no error message, 
when read error information, it will return “+0”, meaning “no error”.   
  If turn off the power supply or send *CLS (clear status) order, the error 
messages in the error queue will be cleared. *RST order will not clear the 
error messages in the error queue. 
Order syntax: 
SYSTem:ERRor? 
Parameter: 
None 
SYSTem:REMote 
This order is used to set the power supply to remote control mode through the 
RS232 interface. Except for the Local key on the front panel, other keys are 
locked and can’t be used. Send or receive orders without first sending the order 
to configure the remote control may cause unpredictable results. 
Order syntax: 
SYST:REM 
Parameter: 
None 
Quest syntax: 
None 
SYSTem:LOCal 
This order is used to set the power supply to panel control mode through the 