Current Control Commands 
  Copyright ©ITECH Electronic Co., Ltd.  28 
[SOURce:]CURRent:PROTection:TRIPed? 
This command is used to query the execution status of overcurrent protection. 
If "1" is returned, it indicates that the overcurrent protection circuit is triggered 
and the OCP status is not cleared. If "0" is returned, the OCP circuit is not 
triggered. 
Order syntax: 
[SOURce:]CURRent:PROTection:TRIPed? 
Return parameters: 
0|1 
[SOURce:]CURRent:PROTection:CLEar 
This command is used to clear the overcurrent protection status. After 
executing this command, the output current will return to the output state before 
the OCP protection occurs and the upper limit current value of the OCP 
overcurrent protection will remain at the previously set value. Please note that 
the external factor causing overcurrent protection should be excluded before 
sending this command. 
Order syntax: 
[SOURce:]CURRent:PROTection:CLEar 
 
 