Current Control Commands 
  Copyright ©ITECH Electronic Co., Ltd.  25 
Chapter7 Current Control Commands 
CURRent {<current 
value>|MINimum|MAXimum|UP|DOWN|DEF} 
This order is used to set the output current. No matter what the current range 
you select currently, the output current is subject to the newest setting value. 
Besides, MIN and MAX can be used as parameters. MIN represents 0A,MAX 
represents the max value of the current range. You may send CURR? MIN or 
CURR? MAX orders to quire the minimum and maximum current within the 
range. DEF represents default value, it also can be used to set value. 
We can also send CURR UP and CURR DOWN orders to increase or decrease 
current, the stepping value should be set with order ” CURR:STEP”. If adjusted 
value has exceeded the current range, then error information will be 
return:error-222. 
Oder syntax: 
[SOUR:]CURR[:LEVel][:IMMediate][:AMPLitude] <NRf> 
Parameters: 
MIN to MAX 
Unit: 
A 
*RST value: 
MIN 
Query syntax: 
CURRent? [MINimum|MAXimum] 
Return parameters: 
<NR2> 
Example: 
CURR:STEP 0.01              //set the stepping value to be 0.01A 
CURR UP              //enable the output current increase value one time 
CURR:STEP 0.02              // set the stepping value to be 0.02A 
CURR DOWN          // enable the output current decrease value one time 