9. Product Specifications
9.1 Delivery form
– Programat EP 3010
– Power cord
– Vacuum hose
– Programat Firing Tray Kit 2
– Automatic Temperature Checking Set ATK2 (test set)
– USB download cable
– Cooling grid
– Various accessories
Recommended accessories
– Automatic Temperature Checking Set ATK2
– Vacuum pump VP5
9.2 Technical data
Power supply 110 – 120 V / 50 – 60 Hz
200 – 240 V / 50 – 60 Hz
Overvoltage category II
Pollution degree 2
Tolerated voltage fluctuations ± 10%
Max. power consumption 12 A at 110 – 120 V
8 A at 200 – 240 V
Acceptable data for pumps of other manufacturers:
Max. output:
Final vacuum:
250 W / max. leakage current 0.75 mA
< 50 mbar
Only tested pumps should be used.
Values of electric fuses
110 – 120 V:
125 V / T15A (heating circuit)
125 V / T15A (vacuum pump)
200 – 240 V:
250 V / T8A (heating circuit)
250 V / T3.15A (vacuum pump)
Dimensions of electric fuses 110 – 120 V: Diameter 6.3 x 32 mm
200 – 240 V: Diameter 5 x 20 mm
Dimensions of the closed furnace Depth: 465 mm
Width: 320 mm / 390 mm (with cooling tray)
Height: 550 mm
Usable size of the firing chamber Diameter: 90 mm
Height: 80 mm
Max. firing temperature 1200 °C
Weight 18.3 kg