SPECIFICATIONS
Model 3000T Operation Manual
36
Pacer Pulse Rejection
Width: 0.1 to 2 ms at ±2 to ±700 mV
Overshoot: Between 4 to 100ms and not greater than 2mV.
Fast ECG signals: 2mV/100µs.
Detector disabling: None.
NOTE: Pacemaker pulses are not present in any rear panel outputs.
Alarms
High Rate: 15 to 250 bpm in 5 bpm increments
Low Rate: 10 to 245 bpm in 5 bpm increments
Asystole: R to R interval >6 seconds
Lead Off: Detached lead or offset potential >0.5 V
Simulator Option
ECG waveform amplitude: 1mV
Simulator rate: Variable rate in steps of 40, 60, 90, 120 and 150 BPM
Also, manually adjustable in increments of 1 BPM.
Test Mode
ECG: 1 mV/100 ms @ 70 bpm
Display
Type: Active Matrix TFT Color LCD (640x480)
Trace: Single ECG trace with “freeze” function.
Screen Size: 13.25cm x 9.94cm, 16.5cm (6.5in) diagonal
Sweep Speed: 25, 50 mm/s
Aspect ratio: 0.4 (standard). User selectable.
USB Port and Data Transfer
Type: Industry standard USB Flash Drive (memory stick) minimum capacity of 512
MB
ECG storage: 100 most recent events (100 high resolution and 100 low resolution)
Impedance Values Storage: 100 most recent events
Mechanical
Size: Height: 6.70in. (17.2cm)
Width: 9.25in. (33.5cm)
Depth: 9.21in. (23.4cm)
Weight: 6.5lbs (2.9kg)
Recorder
Writing Method: Direct Thermal
Number of Traces: 1
Modes: Direct - Manual Recording
Timed - Print button initiates a 30 second recording
Delay - Records 20 seconds before and 20 seconds after an alarm
Paper Speeds: 25 and 50 mm/s
Resolution: Vertical - 200 dots/in.
Horizontal - 600 dots/in. at 25 mm/s
400 dots/in. at >25 mm/s
Frequency Response: >100 Hz at 50 mm/s
Data Rate: 400 samples/s/trace