EasyManuals Logo

Ivy Biomedical Systems 3000T User Manual

Default Icon
48 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #47 background imageLoading...
Page #47 background image
SPECIFICATIONS
Model 3000T Operation Manual
37
Synchronized Output (Trigger)
Test input signal at ECG leads: ½ sine wave, 60ms width, 1mV amplitude, 1 pulse per second
Output trigger delay: < 6ms w/ notch filter off. < 9ms w/ notch filter on
R to R Trigger Accuracy: 50s typical @ 1 mV input
Pulse width: 100ms (50ms available for monitors sold in Japan)
Pulse amplitude: 0 to +5V
Output Impedance: <100
Sensitivity and Threshold
Adjustment: Fully Automatic
Real Time Clock
Resolution: 1 minute
Display: 24 hours
Power Requirement: The real time clock keeps time if the monitor has power or not.
The clock is powered by a dedicated battery whose life is a minimum
4 years at a temperature of 25
o
C
Environmental
Operating Temperature
Range: 5C to 40C
Storage Temperature
Range: -20C to 49C
Relative Humidity: 0-90% non-condensing
Atmospheric Pressure: 500-1060 mbar
Protection against ingress of fluids: IPX0 – Ordinary (without protection against ingress of water)
Power Requirements
Voltage Input: 100 to 230V~
Line Frequency: 47 to 63 Hz
Fuses Type and Rating: T.5A, 250V (Metric 5x20mm)
Maximum ac Power
Consumption: 35 VA
Regulatory
Unit meets or exceeds the specifications for the AAMI Cardiac Monitor Standard EC-13, UL2601-1,
CAN/CSA C22.2 No 601.1-M90, CDN MDR (CMDCAS), IEC 60601-2-25, IEC 60601-2-27,
MDD.93/42/EEC, CE 0143, ISO 13485, and FDA/CGMP.