SPECIFICATIONS
Model 3000T Operation Manual
37
Synchronized Output (Trigger)
Test input signal at ECG leads: ½ sine wave, 60ms width, 1mV amplitude, 1 pulse per second
Output trigger delay: < 6ms w/ notch filter off. < 9ms w/ notch filter on
R to R Trigger Accuracy: 50s typical @ 1 mV input
Pulse width: 100ms (50ms available for monitors sold in Japan)
Pulse amplitude: 0 to +5V
Output Impedance: <100 Ω
Sensitivity and Threshold
Adjustment: Fully Automatic
Real Time Clock
Resolution: 1 minute
Display: 24 hours
Power Requirement: The real time clock keeps time if the monitor has power or not.
The clock is powered by a dedicated battery whose life is a minimum
4 years at a temperature of 25
o
C
Environmental
Operating Temperature
Range: 5C to 40C
Storage Temperature
Range: -20C to 49C
Relative Humidity: 0-90% non-condensing
Atmospheric Pressure: 500-1060 mbar
Protection against ingress of fluids: IPX0 – Ordinary (without protection against ingress of water)
Power Requirements
Voltage Input: 100 to 230V~
Line Frequency: 47 to 63 Hz
Fuses Type and Rating: T.5A, 250V (Metric 5x20mm)
Maximum ac Power
Consumption: 35 VA
Regulatory
Unit meets or exceeds the specifications for the AAMI Cardiac Monitor Standard EC-13, UL2601-1,
CAN/CSA C22.2 No 601.1-M90, CDN MDR (CMDCAS), IEC 60601-2-25, IEC 60601-2-27,
MDD.93/42/EEC, CE 0143, ISO 13485, and FDA/CGMP.