106 – Operating and Maintenance Instructions
Peak Peak
Wavelength
(nm)
Wavelength
Reading (nm)
Difference
within
±0.003 nm?
Peak Peak
Wavelength
(nm)
Wavelength
Reading (nm)
Difference
within
±0.003 nm?
16 1516.8754 10 1530.9770
15 1517.3152 11 1531.5886
14 1517.7613 12 1532.2067
13 1518.2138 13 1532.8312
12 1518.6725 14 1533.4621
11 1519.1376 15 1534.0995
10 1519.6090 16 1534.7433
9 1520.0867 17 1535.3935
8 1520.5707 18 1536.0502
7 1521.0611 19 1536.7134
6 1521.5579 20 1537.3830
5 1522.0610 21 1538.0590
4 1522.5704 22 1538.7416
3 1523.0862 23 1539.4306
2 1523.6084 24 1540.1261
1 1524.1369 25 1540.8281
Figure 89: Absorption Lines of
13
C
2
H
2
Acetylene Cell