108 – Operating and Maintenance Instructions
Peak Peak
Wavelength
(nm)
Wavelength
Reading (nm)
Difference
within ±0.003
nm?
Peak Peak
Wavelength
(nm)
Wavelength
Reading (nm)
Difference
within ±0.003
nm?
28 1550.1776
Calibrate the SWS to the acetylene trace.
1. The objective of calibrating the SWS is to get a best-fit match between the gas cell reference
values and the acquired trace values.
2. On the plot, double click at The first peak position to be recalibrated. The wavelength will then
be shown in the Peak 1 field in the Acquired Acetylene Cell Peaks area of the window.
Double click at a second peak position to be recalibrated. The second wavelength will then
be shown in the Peak 2 field.
3. Click the Re-Calibrate button. The user is asked to continue or advised that the calibration is
complete. Click the Yes button to continue, the software acquires another trace. Double-click
on the two peaks as in the previous step and press recalibrate. Calibration is now complete,
click OK.
4. If step 3 is necessary for more than 5 iterations, a system error is indicated, please contact
JDS Uniphase customer support.
Save the calibration values to the Eeprom in the Control Module in the receiver
To save the calibration values to the EEPROM in the receiver, click on the Save Calibration
button. A dialog box will appear asking if you want to save the previously stored values in the
eeprom to the INI file. Say yes to this option if you plan on reverting to the old values in the future.
Figure 90
To revert to the old calibration, click the Restore Calibration button. The following dialog box will
appear.