Operating and Maintenance Instructions – 113
Use the following table to record the wavelengths of the absorption lines when using the H
13
C
14
N
acetylene cell for calibration.
Table 16: Wavelengths of Absorption Lines for H
13
C
14
N Cell
Peak Center
Wavelength
(nm)
Wavelength
Reading (nm)
Difference
within
±0.003 nm?
Peak Center
Wavelength
(nm)
Wavelength
Reading (nm)
Difference
within
±0.003 nm?
R 25 1528.054 P 1 1543.1148
24 1528.486 2 1543.809
23 1528.927 3 1544.515
22 1529.376 4 1545.231
21 1529.838 5 1545.956
20 1530.306 6 1546.690
19 1530.786 7 1547.435
18 1531.276 8 1548.190
17 1531.774 9 1548.955
16 1532.283 10 1549.730
15 1532.802 11 1550.515
14 1533.329 12 1551.311
13 1533.867 13 1552.116
12 1534.416 14 1552.931
11 1534.972 15 1553.756
10 1535.540 16 1554.589
9 1536.117 17 1555.435
8 1536.703 18 1556.292
7 1537.300 19 1557.157
6 1537.907 20 1558.033
5 1538.522 21 1558.919
4 1539.149 22 1559.814
3 1539.786 23 1560.719
2 1540.431 24 1561.634
1 1541.087 25 1562.563
0 1541.753