116 – Operating and Maintenance Instructions
Figure 95: Power Calibration Window
2. Type the names of the equipment being used for calibration in the Laser Source and Power
Meter fields.
3. If necessary, click the Refresh Channels button to redisplay the channels in the Available
Channels column.
4. Select the Available Channels to be calibrated.
5. Click the > or >> button to load the channels. The channels appear in the Selected Channels
column.
6. Set up the equipment as for power verification, previously shown in Figure 94. The example
shown uses Channel 2 of the receiver.
7. Turn on the second laser source. Ensure the laser power is between 0 and -10 dBm. A laser
wavelength setting of 1550 nm is recommended. Use of a wavelength meter to monitor laser
wavelength is optional.
8. Set the two-output channel HA9 attenuator so that the laser power to the detector module is
between 0 and -40 dBm. This power is monitored using the power meter.
9. Use the software to click the Process button to initiate the calibration process. The current
channel is shown in the Current Channel field, with the Calibration Date.