EasyManua.ls Logo

Julabo MC - Page 11

Julabo MC
58 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
MC
11
F26-MC F32-MC
Working temperature range °C -28 ... 200 -35 ... 200
Temperature stability °C ±0,01 ±0,01
Temperature selection digital
via keypad indication on MULTI-DISPLAY(LED)
remote control via personal computer
indication on monitor
Temperature indication MULTI-DISPLAY (LED)
Resolution °C 0.1
Absolute Temperature Calibration °C ±3
Temperature control PID2
Heater wattage (at 230 V) kW 2,0 2,0
Heater wattage (at 115 V) kW 1,0 1,0
Cooling capacity
Medium ethanol
°C
kW
+20 0 -20
0.26 0.2 0.06
+20 0 -20
0.45 0.39 0.15
Refrigerant R134a R134a
Electronically adj. pump capacity stages 1 ... 4
Flow rate at 0 bar l/min 11 ... 16
Max. pressure at 0 liters bar 0.23 ... 0.45
Electrical connections:
External alarm output 24-0 Vdc / max. 25 mA
Computer interface RS232
Overall dimensions (WxDxH) cm 42x42x42 31x42x64
Bath opening (WxL) cm 12x14 18x12
Bath depth cm 14 15
Filling volume from ... to liters 3 ... 4,5 5,5 ... 8
Weight kg 31 37
Ambient temperature °C 5 ... 40 5 ... 40
Mains power connection 230 V/50 Hz V/ Hz 207-253 / 50 207-253 / 50
Current draw (at 230 V) A 12 13
Mains power connection 230 V/60 Hz V/ Hz ----- 207-253 / 60
Current draw (at 230 V) A ----- 10
Mains power connection 115 V/60 Hz V/ Hz ----- 103-127 / 60
Current draw (at 115 V) A ----- 14
Mains power connection 100 V/50-60 Hz V/ Hz ----- 90-115 / 50-60
Current draw (at 100 V) A ----- 14
All measurements have been carried out at:
rated voltage and frequency ambient temperature: 20 °C
Technical changes without prior notification reserved.

Table of Contents