Operator responsibility – Safety recommendations
10
2.4. Technical specifications
F12-MC F25-MC
Working temperature range °C -20 ... 200 -28 ... 200
Temperature stability °C ±0,01 ±0,01
Temperature selection digital
via keypad indication on MULTI-DISPLAY(LED)
remote control via personal computer
indication on monitor
Temperature indication MULTI-DISPLAY (LED)
Resolution °C 0.1
Absolute Temperature Calibration °C ±3
Temperature control PID2
Heater wattage (at 230 V) kW 2,0 2,0
Heater wattage (at 115 V) kW 1,0 1,0
Cooling capacity
Medium ethanol
°C
kW
+20 0 -20
0.16 0.1 0.02
+20 0 -20
0.26 0.2 0.06
Refrigerant R134a R134a
Electronically adj. pump capacity stages 1 ... 4
Flow rate at 0 bar l/min 11 ... 16
Max. pressure at 0 liters bar 0.23 ... 0.45
Electrical connections:
External alarm output 24-0 Vdc / max. 25 mA
Computer interface RS232
Overall dimensions (WxDxH) cm 20x36x56 23x42x61
Bath opening (WxL) cm 13x15 12x14
Bath depth cm 13 14
Filling volume from ... to liters 3 ... 4,5 3 ... 4,5
Weight kg 23 31
Ambient temperature °C 5 ... 40 5 ... 40
Mains power connection 230 V/50 Hz V/ Hz 207-253 / 50 207-253 / 50
Current draw (at 230 V) A 11 12
Current draw (at 230 V) CH A <9 + 1> <9 + 2>
Mains power connection 230 V/60 Hz V/ Hz 207-253 / 60 207-253 / 60
Current draw (at 230 V) A 11 12
Mains power connection 115 V/60 Hz V/ Hz 103-127 / 60 103-127 / 60
Current draw (at 115 V) A 12 13
Mains power connection 100 V/50-60 Hz V/ Hz 90-120 / 50-60 90-115 / 50-60
Current draw (at 100 V) A 15 13
All measurements have been carried out at:
rated voltage and frequency ambient temperature: 20 °C
Technical changes without prior notification reserved.