Safety installations, warning functions
36
L The warning functions are only triggered when the actual
bath temperature, after start from the „OFF“ or „rOFF“
mode, lies within the set limits for 3 seconds.
Recommendation::
Set the high temperature warning value >t High< 5 °C
to10 °C above the working temperature setpoint
Set the low temperature warning value >t Low< 5 °C to
10 °C below the working temperature setpoint
8.3. Limitation of the operating temperature range
Factory settings:
t HighSetpoint: 200 °C
t LowSetpoint: -50 °C
The limitation of the operating temperature range effects the
temperature setting under the menu called up with the
key. It is possible to adjust only working temperatures that lie
within the limit range set here.
Existing settings for Setpoint 1, 2, 3 and also for > t High <
and > t Low < (see page 35) are automatically defered within
the limit range.
1. Press the key until the menu item
or is displayed.
2. Press enter to indicate the adjusted value on the
MULTI-DISPLAY (LED) (digit blinks).
3. Set value:
Use the cursor keys
to move left or right on the
display until the numeral you wish to change blinks.
Use the increase/decrease arrows
to change the
selected numeral (-, 0, 1, 2, 3, ... 9).
4. Press enter to store the selected value.
Press
to update the display immediately.