10. Troubleshooting guide / Error messages
+8
Alarm with a complete shutdown of the unit
Whenever the microprocessor electronics registers a failure, a
complete shutdown of the heater and circulating pump is
performed. The alarm light „
“ illuminates and a continuous signal
tone sounds. The MULTI-DISPLAY (LED) indicates the cause for
the alarm in form of a code.
Warning without a complete shutdown of the unit
The MULTI-DISPLAY (LED) indicates the cause for the warning in
form of a code and an acoustic signal sounds in regular intervals.
These messages appear every 10 seconds.
Press enter
to quit the audible signal.
The circulator is operated without bath fluid, or the liquid level is
insufficient. Replenish the bath tank with the bath fluid.
Tube breakage has occured (insufficient filling level due to
excessive bath fluid pumped out). Replace the tubing and
replenish the bath tank with the bath fluid.
The float is defect (e. g., because damaged in transit). Repair
by authorized JULABO service personnel.
During the self-test after switch-on a short-circuit is registered
between pin 2 and pin 4 of the control cable or the control cable
is interrupted during operation.
Reconnect the cable or eliminate the short-circuit.
Excess temperature warning or
Excess temperature alarm with a complete shutdown of the
main functional elements being effected.
Warn type: >Li 0 = Warning< or > Li 1 = Alarm<
Low temperature warning or
Low temperature alarm with a complete shutdown of the main
functional elements being effected.
Warn type: >Li 0 = Warning< or > Li 1 = Alarm<
Cable of the working temperature sensor interrupted or short-
circuited.