FP40-MC
Working temperature range °C -40 ... 200
Temperature stability °C ±0,01
Temperature selection digital
via keypad indication on MULTI-DISPLAY(LED)
remote control via personal computer
indication on monitor
Temperature indication MULTI-DISPLAY (LED)
Resolution °C 0.1
Absolute Temperature Calibration °C ±3
Temperature control PID2
Heater wattage (at 230 V) kW 2,0
Heater wattage (at 115 V) kW 1,0
Cooling capacity
Medium ethanol
°C
kW
+20 0 -20 -40
0.68 0.5 0.32 0.04
Refrigerant R404A
Electronically adj. pump capacity stages 1 ... 4
Flow rate at 0 bar l/min 11 ... 16
Max. pressure at 0 liters bar 0.23 ... 0.45
Electrical connections:
External alarm output 24-0 Vdc / max. 25 mA
Computer interface RS232
Overall dimensions (WxDxH) cm 37x46x69
Bath opening (WxL) cm 23x14
Bath depth cm 20
Filling volume from ... to liters 9 ... 16
Weight kg 48
Ambient temperature °C 5 ... 40
Mains power connection 230 V/50 Hz V/ Hz 207-253 / 50
Current draw (at 230 V) A 13
Current draw (at 230 V) (CH) A 9+3
Mains power connection 230 V/60 Hz V/ Hz 207-253 / 60
Current draw (at 230 V) A 13