Recommended quality of cooling water:
pH – value 7,5 to 9,0
Sulfate [SO4 2- ] < 100 ppm
Hydrocarbonate [HCO3- ] / Sulphate [SO4 2-] > 1 ppm
Hardness [Ca2+, Mg2+] / [HCO3-] > 0,5 dH
Alkalinity 60 ppm < [HCO3-] < 300 ppm
Conductivity < 500 μs / cm
Chloride (CL-) < 50 ppm
Phosphate (PO43-) < 2 ppm
Ammonia (NH3) < 0,5 ppm
Free Chlorine < 0,5 ppm
Ferri Ions (Fe3+ ) < 0,5 ppm
Mangano Ions (Mn2+) < 0,05 ppm
Carbon dioxide (CO2) < 10 ppm
Hydrosulfide (H2S) < 50 ppm
Content of oxygen < 0,1 ppm
Algae growth impermissible
Suspended solids impermissible
Notice:
Danger of corrosion of heat exchanger due to unsuitable quality of
cooling water.
Due to its high content of lime hart water is not suitable for cooling
and causes calcination of the heat exchanger.
Ferrous water or water containing ferrous particles will cause
formation of rust even in heat exchangers made of stainless steel.
Chlorous water will cause pitting corrosion in heat exchangers made
of stainless steel.
Due to its corrosive characteristics distilled and deionized water is
unsuitable and will cause corrosion of the bath. .
Due to its corrosive characteristics sea water is not suitable.
Due to its microbiological (bacteria) components which settle in the
heat exchanger untreated and unpurified river water and water from
cooling towers is unsuitable.
Avoid particulate matter in cooling water.
Avoid putrid water.
Notice: Cooling water circuit
Risk of oil leaking from the cooling circuit (compressor) of the recirculating
cooler into the cooling water in case of a fault in the circuit!
Observe the laws and regulations of the water distribution company valid in the
location where the unit is operated.