MC
49
The early warning system for low level signals a critical fluid
level.
Replenish the bath tank with the bath fluid.
After eliminating the malfunction, press the mains power
switch off and on again to cancel the alarm state.
If the unit cannot be returned to operation, contact an authorized
JULABO service station.
Special message „CONFIGURATION ERROR“
The configuration of the circulator does not conform to its present
use.
Press enter
to automatically perform a single modification of
the configuration.
Then contact an authorized JULABO service station.
Disturbances that are not indicated.
The electronic pump motor is overload-protected by an electronic
current limiter. If viscosity of the bath fluid is or becomes too high,
the motor stops running.
15
A
M
P
Mains circuit breakers (resettable) 15 A
Cooling machine: Fuse T 10.0 A, dia.5 x 20 mm
The mains fuses (8b) on the rear of the unit may easily be
exchanged as shown on the left.
Warning:
Before exchanging the fuses, turn off the mains power switch and
disconnect the power plug from the mains socket!
Only use fine fuses with a nominal value as specified.
Example:
Manufacturer Supplier Type Order No.
Wickmann Wickmann G- fuse insert
T10,0A 5x20 mm
No. 19195