EasyManua.ls Logo

JUMO TDA-3000 - 8.7 Digital Filter Configuration

JUMO TDA-3000
68 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
9 Evaluating data with a PC
56
SETUP_*.CSV The file contains the user names and configuration data. As standard, it is
available in the write/read mode and can be edited directly on the TDA.
The file contains the following data:
H
Writing directly to the file on the TDA can be prevented through the
parameter (2) “Lock menu Fn3, Fn8 and Fn9”. If parameter (2) is
set, the file receives the “Read only” attribute.
v See “Lock” on Page 51.
Name Display Value range factory-
set
User name User= 11 characters
v Chapter 4.1
not
specified
Display language Language= 0: German
1: English
2: French
0
Date format Date_format= 0: month/day
1: day/month
1
Sensor type Input= 0: RTD (Pt100)
1: Thermocouple K
2: Thermocouple J
3: Thermocouple T
0
Temperature unit Unit= 0: °C
1: °F
0
Decimal place TempDP= 0: No decimal
place
1: One decimal
place
1
Temperature
correction
TBias= -99.9 to +99.9°C (°F) 0.0
Digital filter TFilter= 0 to 100 seconds
(0: inactive)
0
Switch-off AutoOFF= 0 to 60 minutes
(0: inactive)
3
Interval Interval= 0 to 3600 seconds
(0: manual data
logging)
0

Table of Contents