TECHNICAL DESCRIPTION MULTICAL® 21 & flowIQ® 2101 /..02 /..03
54 Kamstrup A/S •Technical Description • 5512-897_J1_GB • 06.2016
11.8.3 Reset
The meter reset is possible via the optical eye used in connection with re-verification by authorized
laboratories. All loggers and registers, including the legal volume register, will be reset. Only the adjust log
and the clock are not reset. At the same time, the meter is set to transport mode and the radio switches
off. The following restrictions apply to reset:
- The meter can be reset maximum 9 times.
- Each 'reset is logged in the meter.
IMPORTANT: The laboratory, which resets the meter, must attach an adjustment label with the valid
adjustment number to the meter. Naturally, this number must be identical with the number (AX) in the
meter’s display.
Below, samples on:
Re-verification label
Label used at control testing
(meters not approved for re-verification)