12 KANE 458s MANUAL
USING YOUR KANE458s FOR THE FIRST TIME
Charge your analyser’s batteries for 8 hours – an overnight charge
should be sucient for an average 8-hour day.
Take time to read this manual fully & be aware your analyser’s
conguration may not support all features detailed in this manual.
Take time to set up your requirements before using your analyser.
USING YOUR KANE458s EVERY TIME:
Check your analyser’s water trap is empty & particle lter is not dirty.
To empty, pull out the red plug, shake out water & replace red plug.
To change the particle lter, remove protective rubber cover, slide the
water trap from analyser, remove particle lter from its spigot &
replace.
Reconnect your analyser’s water trap & rubber protective cover.
Connect your ue probe hose to your analyser’s ue gas inlet – see p.8.
Connect your ue probes temperature plug to your analyser’s T1 socket
– check plug orientation is correct – See p.8.
FRESH AIR PURGE ON SWITCH ON
Your analyser starts a zero countdown to auto calibrate itself.
When nished, rotate dial to “RATIO” – In fresh air CO reading should be
zero.
Rotate dial to “O2/EFF” – In fresh air O2 readings should be 20.9% +
0.3%.