29KANE 458s MANUAL
ANALYSER PROBLEM SOLVING
FAULT SYMPTOM CAUSES / SOLUTIONS
• Oxygen too high
• CO2 too low
Air leaking into probe, tubing, water trap,
connectors or internal to analyser
• Batteries not holding charge
Analyser not running on mains adapter
• Batteries exhausted
AC charger not giving correct output
• No fuse
•Analyser does not respond to ue gas
• Particle lter blocked
• Probe or tubing blocked
• Pump not working or damaged with
contaminants
• Net temperature or Efciency calculation
incorrect
Ambient temperature set wrong during
Automatic Calibration
•Flue temperature readings erratic
• Probe not connected
• Faulty connection or break in cable or plug
• T ue or ΔT displays (-N/F-)
• Probe not connected.
• Faulty connection or break in cable or plug
• EFF or X-Air displays (- O2++-)
• CO2 reading is below 2%.
• 02 > 18%
Analyser just continually beeps
• Turn dial back to MENU and press
• Turn dial back to Tightness and press