KANE905 Operators Manual Page 24
Page 1:
LANG
: ENGLISH
CO MENU
:
CALENDAR
:
ZERO
: 90
Page 2:
NOx%
: -
HEADER
PRINT
: KANE IRP2
Parameter
Description
Settings
LANG:
Changes the analysers displayed and
printed language.
ENGLISH
SPANISH
DUTCH
FRENCH
ITALIAN
ZERO:
Allows setting of the Autocalibration time
in minutes. Care must be taken when
changing this parameter as sensors may
drift from zero if too short a time is used.
Kane International advise 3 minute
countdown.
90 seconds
120 seconds
180 seconds
300 seconds
NO REF:
Displayed on the Nitric Oxide unit only.
Allows the percentage P in the following
calculation to be set. The default value set
is 5%. Note the percentage allows for
NO
2
in a typical boiler.
NO
x
= NO + P% NO
OFF
1-9%
CALENDAR:
Allows the user to change the date and
time. (24 hour clock).
This screen will be shown once
the parameter is entered:
hh : mm : ss
TIME
13 : 53 : 26
FORMAT
dd : mm : yy
DATE
23 07 : 15