CONTINUED MENU CHOICES
KANE458s MANUAL 19
NOX CALC
NOX CALC
(If NO sensor fitted)
Set REFERENCE NOx to percentage
required or defined by local regulations.
It is typically assumed boiler emissions of
NOx = NO + 5% NO2
If required, you can calculate NOx from 3
options - SUM, NO or NO2.
SUM adds NO & NO2 if both sensors are
fitted.
NO calculates NOx from NO, if fitted, where
NOx = NO x 1.1
NO2 calculates NOx from NO2 only, if fitted,
where NOx = NO2 x 2.05.
Use or to select then press ENTER
LOGS LOGS View current memory usage & stored logs
EFF
EFF
Select EFF to set efficiency scale by
pressing ENTER.
Select NETT or GROSS using & , then
press ENTER to confirm
Considering calculation automatically
selected based on selected fuel
UTIL
UTIL
Select UTIL to access utility menu - Press
ENTER then or & ENTER to select
from:
INFO - Shows Analyser firmware, wireless
firmware & date next calibration due
LEAK - Performs system leak test following
onscreen instructions
B’LIGHT - Selects time backlight is on from
15 to 300 seconds using or & ENTER
to confirm
NOTE: Extending backlight time reduces
analyser battery life
CODE
CODE
Password protected for authorised service
agents only - Default to 000000