23
ABSORmatic
®
plus.
9.10 Assist programm “memorisation of
the manual sensibilitity adjustment“
(2350)
By pressing foil switch “up“ or “down“
call up the assist programm “manual
sensib.“and press foil switch enter .
Select the desired workstation
(WP1/WP2/WP3) by pressing foil switch
“up“ or “down“
Press foil switch enter . By pressing foil
switch “up“ or “down“ the memo-
rised setting altered.
Press foil switch . The new setting will
be programmed
By pressing foil switch “back“ the assist
programm can be left again.
Factory setting: 116 (230 V)
260 (100 V)
200 (120 V)
9.11 Assist programm: “Display of the
socket current“ (2400)
By pressing foil switch “up“ or “down“
call up assistprogramm “socket current“
and press foil switch enter .
Select the desirede work station
(WP1/WP2/WP3) by pressing foil switch
“up“ or “down“
Press foil switch enter . The, exactly at
this point in time, measured socket current
is displayed. This readout is without a spe-
cific measuring unit. The readout fits
though directly to the assist programms
“manual sensibility adjustement“ and
“automatic sensibility adjustement“
With foil switch “left-back“ one can
leave again the assist programm “socket
current“
Û
Á
ˇ
˝
(2350) 9.10 Manual gain setting: 2351 WP1 0..1023
2352 WP2 0..1023
2353 WP1 0..1023
2354 WP21 0..1023
2355 WP22 0..1023
2356 WP23 0..1023
2357 WP24 0..1023
(2400) 9.11 Display socket currents: 2401 SO1
2402 SO2
2403 SO3
2404 SO21
2405 SO22
2406 SO23
2407 SO24