11.9 Error message:
Warning 32 - (1201) Service life AKF
(active carbon filter)
I• Memory to check or exchange the car-
bon filter .
C• The obervation periode for the carbon
filter is expired.
R• Indication "Filter ok" prolongate the obs-
ervation periode.
Indication "Filter worn-out" exchange
active carbon filter.
11.10 Error message:
I• Suction appears only at one/two
instead of three suction points.
C• large foreign body sucked in
R• Check suction line for blockage and
remove foreign body. Clean suction line.
C• dirt grid, fine gold filter blocked
R• Clean dirt grid and possible exchange
fine gold filter.
11.11 Error message:
I• Suction present at all 3 suction points
although only one suction passage
needed.h.
C• No compressed air supply at suction
unit diverter .
R• Provide compressed air supply and
check air pressure and if necessary
reset.
11.12 Fehleranzeige:
I• Suction power too low.
C• Suction line(s) blocked.
R• Check suction lines for blockage and
clean.
C• Dirt grid, fine gold filter blocked.
R• Clean dirt grid; empty, clean or if
necessary exchange fine gold filter.
C• Too low suction stage selected
R• Select next higher suction stage
(with pushbutton ).
I = Instructions C = Cause R = Remedy