Instructions for use
ELECTROmatic M/C and PM/PC

Table of Contents

Questions and Answers

  • J
    jonathan36Sep 3, 2025
    Why is the LUX light not on in my KaVo ELECTROmatic PM Dental equipment?
    • D
      Debbie JonesSep 3, 2025
      The LUX light might not be on because the LED is defective. Replace the LED. Also, the motor hose could be defective, so replace the motor hose. Finally, the device itself could be defective, so replace the control unit. If error persists, notify the service engineer.
  • D
    Douglas HowardSep 2, 2025
    What to do if my KaVo Dental equipment malfunctions (no display, the LED on the rear of the device flashes)?
    • J
      Jordan Thompson MDSep 2, 2025
      If your device malfunctions with no display but the LED on the rear flashes, it could be due to no voltage supply on the control panel or no connection to the control panel. Check and restore the correct connection. Try replacing the connection cord to the control panel or replacing the control panel itself. If error persists, notify the service engineer.
  • K
    Kristen WatersSep 1, 2025
    What to do if my KaVo ELECTROmatic PM Dental equipment malfunctions (no display, the LED on the rear of the device does not flash)?
    • S
      Steven SanfordSep 2, 2025
      If your device malfunctions with no display and the LED on the rear doesn't flash, the cause might be no voltage supply. Check and restore the correct voltage supply and connection. Check that the standby LED on the power supply is on. If the standby LED on the power supply is not on, replace the power supply. If error persists, notify the service engineer.
  • J
    Jennifer MoralesAug 31, 2025
    Why is there no or insufficient spray water in my KaVo ELECTROmatic PM?
    • D
      destiny22Sep 1, 2025
      There could be several reasons for insufficient spray water. The spray water regulation might be closed, so open it as much as possible. The water line could be soiled; in this case, pull off the handpiece and actuate the spray water in bursts to rinse it. Another cause may be that the filter insert is soiled, so replace the filter on the water inlet. Finally, the spray water on the treatment center or foot control could be closed, so open the spray on the treatment center. If the error persists, unscrew the turbine hose and check if any water flows. If no water flows, notify a service engineer.
  • C
    Christopher RobinsonAug 30, 2025
    What to do if SAFEdrive icon flashes yellow on my KaVo Dental equipment?
    • C
      Chloe CarterAug 31, 2025
      If the SAFEdrive icon is flashing yellow, it means the motor is overloaded. If you are using a handpiece from a third-party manufacturer, deactivate SAFEdrive or use a high-speed handpiece from KaVo. If you are using a KaVo high-speed handpiece, relieve the handpiece of its load for at least 2 seconds by lifting it off the tooth. Once the motor reaches maximum speed again and the SAFEdrive icon turns yellow, continue working as usual, making sure to use intermittent mode as the operating mode.
  • N
    Nicole SteeleAug 29, 2025
    What to do if SAFEdrive icon flashes red on my KaVo ELECTROmatic PM?
    • K
      kingdillonAug 30, 2025
      If the SAFEdrive icon flashes red and the motor stops automatically, it indicates the motor is overloaded. If you're using a third-party handpiece, try deactivating SAFEdrive or switch to a KaVo high-speed handpiece. If you're already using a KaVo high-speed handpiece, take it out of the patient's mouth and inspect the handpiece head for temperature, damage, and ease of operation. If there's no damage or overheating, restart the motor by pressing the foot control again. Service the handpiece. However, if you notice damage or overheating, replace the handpiece or have it repaired.
  • S
    Shelly FrankAug 29, 2025
    What does Event E2 mean on KaVo Medical Equipment?
    • J
      Jenna BradyAug 29, 2025
      Event E2 on your KaVo Medical Equipment indicates that the foot control is being pressed during start-up. Avoid pressing the foot control when switching the unit on.
  • W
    William CunninghamAug 28, 2025
    What does Event E3 mean on KaVo Medical Equipment?
    • R
      Roger AtkinsonAug 28, 2025
      Event E3 on your KaVo Medical Equipment means that no motor is connected. Connect the motor.
  • M
    Mrs. Samantha HowardAug 27, 2025
    What does Event E4 mean on KaVo Medical Equipment?
    • R
      robinsondouglasAug 27, 2025
      Event E4 on your KaVo Medical Equipment indicates that the motor is blocked. To resolve this, relieve the motor of its load, then stop and start it using the foot control.
  • L
    Luke PerryAug 26, 2025
    What does Event E5 mean on KaVo Medical Equipment?
    • C
      Cory HerreraAug 26, 2025
      Event E5 on your KaVo Medical Equipment indicates automatic light and motor shut-off during continuous operation of the motor. Comply with the defined operating mode.

Summary

User Instructions

User Guide

General overview and essential information for safe and correct operation of the device.

Safety

Infection Hazard

Identifies risks of infection from contaminated devices and outlines necessary precautions.

Explosion Hazard Area

Warns about risks of explosion or fire due to electrical sparks in specific environments.

Technical Condition

Emphasizes maintaining the product in good technical condition to prevent injury.

Ingress of Liquids

Details risks of electrical shock and injury from liquid ingress and how to prevent it.

Accessories and Combination with Other Equipment

Safety guidelines for using approved accessories and avoiding unauthorized modifications.

Qualification of Personnel

Stresses the need for trained personnel and proper understanding of instructions for safe operation.

Service and Repair

Outlines who is authorized for service/repair and the procedures required for safety checks.

Description of the Product

Intended Use

Defines the product's purpose, indications, and proper usage for dental treatments.

Technical Specifications of the ELECTROmatic

Provides detailed technical data including dimensions, weight, ambient conditions, and electrical ratings.

Installation

Installation Positions

Outlines various methods and positions for physically mounting the ELECTROmatic unit.

Startup

Operation

Starting-up the Motor

Instructions on how to start the motor using the foot control and achieve desired speeds.

Regulating the Spray Water

Explains how to adjust the spray water volume for cooling and lubrication.

Changing the Speed Setting

Describes how to change and save motor speed presets using the control panel or foot control.

Protection Function SAFEdrive (PM;PC Only)

Details the SAFEdrive function for detecting and preventing overheating of handpieces.

Reprocessing Steps in Accordance with DIN EN ISO 17664

Cleaning

Guidelines for cleaning the device, including preparation, external, and internal cleaning procedures.

Disinfection

Instructions for disinfecting the device surfaces using approved methods and products.

Servicing

Troubleshooting

KaVo ELECTROmatic PM Specifications

General IconGeneral
Supply voltage100-240 V
Frequency50/60 Hz
Power60 W
TypeElectric motor
Voltage24 V
Handpiece ConnectionISO 3964