Performance Verification
1-8
2002
ACV
range
Applied
voltage
Reading limits (1 year, 18°C to 28°C)
100Hz 1kHz 5kHz 25kHz 50kHz 100kHz 200kHz 1MHz 2MHz
200mV 190.000mV 189.914mV
to
190.087mV
189.942mV
to
190.058mV
189.942mV
to
190.058mV
189.933mV
to
190.068mV
189.885mV
to
190.115mV
189.400mV
to
190.600mV
188.525mV
to
191.475mV
186.000mV
to
194.000mV
180.100mV
to
199.900mV
2V 1.90000V 1.89914V
to
1.90087V
1.89942V
to
1.90058V
1.89942V
to
1.90058V
1.89933V
to
1.90068V
1.89885V
to
1.90115V
1.89400V
to
1.90600V
1.88525V
to
1.91475V
1.86000V
to
1.94000V
1.80100V
to
1.99900V
20V 19.0000V 18.9885V
to
19.0116V
18.9913V
to
19.0087V
18.9894V
to
19.0106V
18.9875V
to
19.0125V
18.9837V
to
19.0163V
18.9400V
to
19.0600V
18.8525V
to
19.1475V
18.2000V
to
19.8000V
*
200V 190.000V 189.878V
to
190.122V
189.906V
to
190.094V
189.887V
to
190.113V
189.868V
to
190.132V
189.830V
to
190.170V
189.393V
to
190.607V
***
750V 750.00V 748.98V
to
751.02V
749.09V
to
750.91V
749.02V
to
750.98V
748.87V
to
751.13V
*****
** CAUTION: Do not exceed 2 × 10
7
V•Hz input.
** Use wideband option and connections for 2MHz tests.
NOTE: Reading limits shown include total absolute uncertainty of recommended calibrator (see Table 1-1). Reading limits also include the adder for AC Coupling of the input.
Table 1-3
Limits for normal mode AC voltage verification

Table of Contents

Other manuals for Keithley 2002

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2002 and is the answer not in the manual?

Keithley 2002 Specifications

General IconGeneral
Resolution8.5 digits
DC Voltage Accuracy±0.0006%
Maximum DC Voltage1000 V
Maximum AC Voltage750 V
Maximum DC Current3 A
Maximum AC Current3 A
ConnectivityGPIB, RS-232
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Resistance Range10 Ω to 1 GΩ
Model2002
CategoryMultimeter
InterfaceGPIB, RS-232
AC Current Accuracy±0.1%
Frequency Range0.1Hz to 300kHz
DisplayVFD
TypeDigital
Power Supply100-240V AC, 50/60Hz
Current Range (DC)100 μA to 3 A
Current Accuracy (DC)±0.01%
Current Range (AC)100 μA to 3 A
Current Accuracy (AC)±0.1%
DC Current Range100 μA to 3 A
AC Current Range100 μA to 3 A

Related product manuals