Calibration Command Reference
3-10
:A2M (CALibration:PROTected:DC:A2M)
Purpose To program the 2mA comprehensive calibration step.
Format :cal:prot:dc:a2m <Cal_current>
Parameter <Cal_current> = 0.95E-3 to 2.05E-3 [A]
Description :A2M programs the 2mA comprehensive calibration step. The allowable range of the calibration
current parameter is from 0.95mA to 2.05µA. Use the 2mA value whenever possible for best
results.
Example :CAL:PROT:DC:A2M 2E-3 Program 2mA step.
:A20M (CALibration:PROTected:DC:A20M)
Purpose To program the 20mA comprehensive calibration step.
Format :cal:prot:dc:a20m <Cal_current>
Parameter <Cal_current> = 9.5E-3 to 20.5E-3 [A]
Description :A200U programs the 20mA comprehensive calibration step. The allowable range of the cali-
bration current parameter is from 9.5mA to 20.5mA. Use the 20mA value whenever possible for
best results.
Example :CAL:PROT:DC:A20M 20E-3 Program 20mA step.
:A200M (CALibration:PROTected:DC:A200M)
Purpose To program the 200mA comprehensive calibration step.
Format :cal:prot:dc:a200m <Cal_current>
Parameter <Cal_current> = 95E-3 to 205E-3 [A]
Description :A200M programs the 200mA comprehensive calibration step. The allowable range of the cal-
ibration current parameter is from 95mA to 205mA. Use the 200mA value whenever possible
for best results.
Example :CAL:PROT:DC:A200M 200E-3 Program 200mA step.

Table of Contents

Other manuals for Keithley 2002

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2002 and is the answer not in the manual?

Keithley 2002 Specifications

General IconGeneral
Resolution8.5 digits
DC Voltage Accuracy±0.0006%
Maximum DC Voltage1000 V
Maximum AC Voltage750 V
Maximum DC Current3 A
Maximum AC Current3 A
ConnectivityGPIB, RS-232
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Resistance Range10 Ω to 1 GΩ
Model2002
CategoryMultimeter
InterfaceGPIB, RS-232
AC Current Accuracy±0.1%
Frequency Range0.1Hz to 300kHz
DisplayVFD
TypeDigital
Power Supply100-240V AC, 50/60Hz
Current Range (DC)100 μA to 3 A
Current Accuracy (DC)±0.01%
Current Range (AC)100 μA to 3 A
Current Accuracy (AC)±0.1%
DC Current Range100 μA to 3 A
AC Current Range100 μA to 3 A

Related product manuals