Calibration Command Reference
3-9
:OHM200 (CALibration:PROTected:DC:OHM200)
Purpose To program the 200 comprehensive calibration step.
Format :cal:prot:dc:ohm200 <Cal_resistance>
Parameter <Cal_resistance> = 95 to 205 []
Description :OHM200 programs the 200 comprehensive calibration step. The allowable range of the
calibration resistance parameteris from 95 to 205. Use the 200 value whenever possible,
or the closest possible value (for example, 190, which is the closet value available on many
calibrators).
Example :CAL:PROT:DC:OHM200 190 Program 200 step.
:OHM20 (CALibration:PROTected:DC:OHM20)
Purpose To program the 20 comprehensive calibration step.
Format :cal:prot:dc:ohm20 <Cal_resistance>
Parameter <Cal_resistance> = 9.5 to 20.5 []
Description :OHM20 programs the 20 comprehensive calibration step. The allowable range of the
calibration resistance parameter is from 9.5 to 20.5k. Use the 20 value whenever possible,
or the closest possible value (for example, 19, which is the closet value available on many
calibrators).
Example :CAL:PROT:DC:OHM20 19 Program 20 step.
:A200U (CALibration:PROTected:DC:A200U)
Purpose To program the 200µA comprehensive calibration step.
Format :cal:prot:dc:a200u <Cal_current>
Parameter <Cal_current> = 95E-6 to 205E-6 [A]
Description :A200U programs the 200µA comprehensive calibration step. The allowable range of the cali-
bration current parameter is from 95µA to 205µA. Use the 200µA value whenever possible for
best results.
Example :CAL:PROT:DC:A200U 200E-6 Program 200µA step.

Table of Contents

Other manuals for Keithley 2002

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2002 and is the answer not in the manual?

Keithley 2002 Specifications

General IconGeneral
Resolution8.5 digits
DC Voltage Accuracy±0.0006%
Maximum DC Voltage1000 V
Maximum AC Voltage750 V
Maximum DC Current3 A
Maximum AC Current3 A
ConnectivityGPIB, RS-232
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Resistance Range10 Ω to 1 GΩ
Model2002
CategoryMultimeter
InterfaceGPIB, RS-232
AC Current Accuracy±0.1%
Frequency Range0.1Hz to 300kHz
DisplayVFD
TypeDigital
Power Supply100-240V AC, 50/60Hz
Current Range (DC)100 μA to 3 A
Current Accuracy (DC)±0.01%
Current Range (AC)100 μA to 3 A
Current Accuracy (AC)±0.1%
DC Current Range100 μA to 3 A
AC Current Range100 μA to 3 A

Related product manuals