Calibration Command Reference
3-8
:OHM200K (CALibration:PROTected:DC:OHM200K)
Purpose To program the 200k comprehensive calibration step.
Format :cal:prot:dc:ohm200k <Cal_resistance>
Parameter <Cal_resistance> = 95E3 to 205E3 []
Description :OHM200K programs the 200k comprehensive calibration step. The allowable range of the
calibration resistance parameter is from 95k to 205k. Use a 100k value whenever possible,
or the closest possible value.
Example :CAL:PROT:DC:OHM200K 100E3 Program 200k step.
:OHM20K (CALibration:PROTected:DC:OHM20K)
Purpose To program the 20k comprehensive calibration step.
Format :cal:prot:dc:ohm20k <Cal_resistance>
Parameter <Cal_resistance> = 9.5E3 to 20.5E3 []
Description :OHM20K programs the 20k comprehensive calibration step. The allowable range of the cal-
ibration resistance parameter is from 9.5k to 20.5k. Use the 20k value whenever possible,
or the closest possible value (for example, 19k, which is the closet value available on many
calibrators).
Example :CAL:PROT:DC:OHM20K 19E3 Program 20k step.
:OHM2K (CALibration:PROTected:DC:OHM2K)
Purpose To program the 2k comprehensive calibration step.
Format :cal:prot:dc:ohm2k <Cal_resistance>
Parameter <Cal_resistance> = 950 to 2.05E3 []
Description :OHM2K programs the 2k comprehensive calibration step. The allowable range of the cali-
bration resistance parameter is from 0.95k to 2.05k. Use the 2k value whenever possible,
or the closest possible value (for example, 1.9k, which is the closet value available on many
calibrators).
Example :CAL:PROT:DC:OHM2K 1.9E3 Program 2k step.