EasyManua.ls Logo

Keithley 2002

Keithley 2002
114 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Calibration Command Reference
3-7
:V2 (:CALibration:PROTected:DC:V2)
Purpose To program the +2V comprehensive calibration step.
Format :cal:prot:dc:v2 <Cal_voltage>
Parameter <Cal_voltage> = 0.95 to 2.05 [V]
Description :V2 programs the +2V DC comprehensive calibration step. The allowable range of the calibra-
tion voltage parameter is from 0.95 to 2.05V, but 2V is recommended for best results.
Example :CAL:PROT:DC:V2 2 Program 2V step.
:V20 (:CALibration:PROTected:DC:V20)
Purpose To program the +20V DC comprehensive calibration step.
Format :cal:prot:dc:v20 <Cal_voltage>
Parameter <Cal_voltage> = 9.5 to 20.5 [V]
Description :V2 programs the +20V DC comprehensive calibration step. The allowable range of the calibra-
tion voltage parameter is from 9.5 to 20.5V, but 20V is recommended for best results.
Example :CAL:PROT:DC:V20 20 Program 20V step.
:OHM1M (CALibration:PROTected:DC:OHM1M)
Purpose To program the 1M comprehensive calibration step.
Format :cal:prot:dc:ohm1m <Cal_resistance>
Parameter <Cal_resistance> = 475E3 to 1.025E6 []
Description :OHM1M programs the 1M comprehensive calibration step. The allowable range of the cali-
bration resistance parameter is from 475k to 1.025M. Use the 1M value whenever possi-
ble, or the closest possible value.
Example :CAL:PROT:DC:OHM1M 1E6 Program 1M cal step.

Table of Contents

Other manuals for Keithley 2002

Related product manuals