Calibration Command Reference
3-7
:V2 (:CALibration:PROTected:DC:V2)
Purpose To program the +2V comprehensive calibration step.
Format :cal:prot:dc:v2 <Cal_voltage>
Parameter <Cal_voltage> = 0.95 to 2.05 [V]
Description :V2 programs the +2V DC comprehensive calibration step. The allowable range of the calibra-
tion voltage parameter is from 0.95 to 2.05V, but 2V is recommended for best results.
Example :CAL:PROT:DC:V2 2 Program 2V step.
:V20 (:CALibration:PROTected:DC:V20)
Purpose To program the +20V DC comprehensive calibration step.
Format :cal:prot:dc:v20 <Cal_voltage>
Parameter <Cal_voltage> = 9.5 to 20.5 [V]
Description :V2 programs the +20V DC comprehensive calibration step. The allowable range of the calibra-
tion voltage parameter is from 9.5 to 20.5V, but 20V is recommended for best results.
Example :CAL:PROT:DC:V20 20 Program 20V step.
:OHM1M (CALibration:PROTected:DC:OHM1M)
Purpose To program the 1M comprehensive calibration step.
Format :cal:prot:dc:ohm1m <Cal_resistance>
Parameter <Cal_resistance> = 475E3 to 1.025E6 []
Description :OHM1M programs the 1M comprehensive calibration step. The allowable range of the cali-
bration resistance parameter is from 475k to 1.025M. Use the 1M value whenever possi-
ble, or the closest possible value.
Example :CAL:PROT:DC:OHM1M 1E6 Program 1M cal step.

Table of Contents

Other manuals for Keithley 2002

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2002 and is the answer not in the manual?

Keithley 2002 Specifications

General IconGeneral
Resolution8.5 digits
DC Voltage Accuracy±0.0006%
Maximum DC Voltage1000 V
Maximum AC Voltage750 V
Maximum DC Current3 A
Maximum AC Current3 A
ConnectivityGPIB, RS-232
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Resistance Range10 Ω to 1 GΩ
Model2002
CategoryMultimeter
InterfaceGPIB, RS-232
AC Current Accuracy±0.1%
Frequency Range0.1Hz to 300kHz
DisplayVFD
TypeDigital
Power Supply100-240V AC, 50/60Hz
Current Range (DC)100 μA to 3 A
Current Accuracy (DC)±0.01%
Current Range (AC)100 μA to 3 A
Current Accuracy (AC)±0.1%
DC Current Range100 μA to 3 A
AC Current Range100 μA to 3 A

Related product manuals