Section
1: Calibrating and adjusting the Model 2110 Model 2110 5½ Digit Multimeter
Calibration Manual
1-4 2110-905-01 Rev. C / August 2013
Recommended test equipment
Use the Fluke 5700A Calibrator and the Fluke 5725A Amplifier (or the equivalent or better) and the
Fluke 5520A/5522A for temperature and capacitance to calibrate the Model 2110.
Calibration code
To unlock and initiate calibration:
1. Unlock calibration by sending the following command:
CAL:PROT:CODE <up to 8-character string>
The default command is:
CAL:PROT:CODE 123456
2. Initiate calibration by sending the following command:
CAL:PROT:INIT

Other manuals for Keithley 2110

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2110 and is the answer not in the manual?

Keithley 2110 Specifications

General IconGeneral
Model2110
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
ContinuityYes
Diode TestYes
CategoryDigital Multimeter
Display5.5 digit
DC Current Range100 µA to 3 A
AC Current Range100 µA to 3 A
Resistance Range100 MΩ
Frequency Range3 Hz to 300 kHz
Frequency Accuracy0.01% + 1 count
Capacitance Range1 nF to 100 µF
Capacitance Accuracy1% + 5 counts
ConnectivityUSB, GPIB (optional)
Dimensions88 mm
Power Supply100 V to 240 V AC, 50/60 Hz
DC Voltage Accuracy±0.012% + 2 counts