EasyManuals Logo

Keithley 2110 Calibration Manual

Keithley 2110
35 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #27 background imageLoading...
Page #27 background image
Section
2: Adjustment Model 2110 5½ Digit Multimeter
Calibration Manual
2-16 2110-905-01 Rev. C / August 2013
Temperature function adjustment
1. Set the Model 5520A/5522A to 0 Deg C, and the output mode to STANDBY.
2. Using Type K thermocouple wire with TC jacks on both ends, connect the Fluke 5520A/5522A TC
output to the Model 2110 input TC terminals. Allow connections to stabilize for a minimum of five
minutes before proceeding.
3. Configure the Model 2110 for the type K by sending the following command:
CONF:TCOUPLE
4. Set the Model 5520A/5522A output to 0 Deg C, type K Thermocouple, and the output mode to
OPERATE.
5. Adjust the upper region of the range by sending the following command:
CAL:PROT:DC:STEP 1,0
6. Wait until the following message is displayed on the Model 2110:
CALI OK
7. Set the Model 5520A/5522A output mode to STANDBY. Press RESET on the Model
5520A/5522A front panel to reset the calibrator to its default settings.
Capacitance function adjustment
1. Set the Model 5520A/5522A to 1 µF range and the output mode to STANDBY.
2. Using shielded, low thermal EMF cables, connect the Fluke 5700A output HI and LO to the Model
2110 input HI and LO terminals. Allow connections to stabilize for a minimum of five minutes
before proceeding.
3. Configure the Model 2110 for the 1 µF capacitance range by sending the following command:
CONF:CAP 1E-6
4. Set the Model 5520A/5522A output mode to STANDBY.
5. Adjust input offset by sending the following command:
CAL:PROT:DC:STEP 1,0
6. Wait until the following message is displayed on the Model 2110:
Cali OK
7. Set the Model 5520A/5522A output to 1UF.
8. Adjust full scale measurement by sending the following command:
CAL:PROT:DC:STEP 2,1E-6
9. Wait until the following message is displayed on the Model 2110:
Cali OK
10. Configure the Model 2110 for the 10 µF Capacitance range by sending the following command:
CONF:CAP 1E-5
11. Set the Model 5520A/5522A output mode to STANDBY.
12. Adjust input offset by sending the following command:
CAL:PROT:DC:STEP 1,0
13. Wait until the following message is displayed on the Model 2110:
Cali OK

Other manuals for Keithley 2110

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2110 and is the answer not in the manual?

Keithley 2110 Specifications

General IconGeneral
Model2110
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
ContinuityYes
Diode TestYes
CategoryDigital Multimeter
Display5.5 digit
DC Current Range100 µA to 3 A
AC Current Range100 µA to 3 A
Resistance Range100 MΩ
Frequency Range3 Hz to 300 kHz
Frequency Accuracy0.01% + 1 count
Capacitance Range1 nF to 100 µF
Capacitance Accuracy1% + 5 counts
ConnectivityUSB, GPIB (optional)
Dimensions88 mm
Power Supply100 V to 240 V AC, 50/60 Hz
DC Voltage Accuracy±0.012% + 2 counts

Related product manuals