Section
2: Adjustment Model 2110 5½ Digit Multimeter
Calibration Manual
2-4 2110-905-01 Rev. C / August 2013
ACV range adjustment
1. Set the Model 5700A to 0.01 V at 1 kHz and the output mode to STANDBY.
2. Using coaxial cable with dual banana terminals connect the Fluke 5700A output HI and LO to the
Model 2110 input HI and LO terminals. Allow connections to stabilize for a minimum of five
minutes before proceeding.
3. Configure the Model 2110 for the 0.1 V range by sending the following command:
CONF:VOLT:AC 0.1
4. Set the Model 5700A output mode to OPERATE.
5. Adjust to 10 percent of range by sending the following command:
CAL:PROT:AC:STEP 1,0.01
6. Wait until the following message is displayed on the Model 2110:
Cali OK
7. Set the Model 5700A output to 0.1 V.
8. Adjust full scale measurement by sending the following command:
CAL:PROT:AC:STEP 2,0.1
9. Wait until the following message is displayed on the Model 2110:
Cali OK
10. Set the Model 5700A output mode to STANDBY.
11. Configure the Model 2110 for the 1 V range by sending the following command:
CONF:VOLT:AC 1
12. Set the Model 5700A to 0.1 V and the output mode to OPERATE.
13. Adjust to 10 percent of range by sending the following command:
CAL:PROT:AC:STEP 1,0.1
14. Wait until the following message is displayed on the Model 2110:
Cali OK
15. Set the Model 5700A output to 1 V.
16. Adjust full scale measurement by sending the following command:
CAL:PROT:AC:STEP 2,1
17. Wait until the following message is displayed on the Model 2110:
Cali OK
18. Set the Model 5700A output mode to STANDBY.
19. Configure the Model 2110 for the 10 V range by sending the following command:
CONF:VOLT:AC 10
20. Set the Model 5700A to 1 V and the output mode to OPERATE.
21. Adjust to 10 percent of range by sending the following command:
CAL:PROT:AC:STEP 1,1
22. Wait until the following message is displayed on the Model 2110:
Cali OK
23. Set the Model 5700A output to 10 V.
24. Adjust full scale measurement by sending the following command:
CAL:PROT:AC:STEP 2,10
25. Wait until the following message is displayed on the Model 2110:
Cali OK

Other manuals for Keithley 2110

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2110 and is the answer not in the manual?

Keithley 2110 Specifications

General IconGeneral
Model2110
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
ContinuityYes
Diode TestYes
CategoryDigital Multimeter
Display5.5 digit
DC Current Range100 µA to 3 A
AC Current Range100 µA to 3 A
Resistance Range100 MΩ
Frequency Range3 Hz to 300 kHz
Frequency Accuracy0.01% + 1 count
Capacitance Range1 nF to 100 µF
Capacitance Accuracy1% + 5 counts
ConnectivityUSB, GPIB (optional)
Dimensions88 mm
Power Supply100 V to 240 V AC, 50/60 Hz
DC Voltage Accuracy±0.012% + 2 counts

Related product manuals