Section
2: Adjustment Model 2110 5½ Digit Multimeter
Calibration Manual
2-18 2110-905-01 Rev. C / August 2013
Calibration date
1. Update the calibration date by sending the following command:
CAL:PROT:DATE <”MM/DD/YYYY”>
The date <string> must be enclosed in double quotes (“<string>”).
Example: If calibration date is 5 Apr 2012, send the following command:
CAL:PROT:DATE ”04/05/2012”
Calibration due date is automatically calculated for 12 months.
2. Save calibration constants in EEPROM by sending the following command:
CAL:PROT:SAVE
3. Lock calibration by sending the following command:
CAL:PROT:LOCK

Other manuals for Keithley 2110

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2110 and is the answer not in the manual?

Keithley 2110 Specifications

General IconGeneral
Model2110
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
ContinuityYes
Diode TestYes
CategoryDigital Multimeter
Display5.5 digit
DC Current Range100 µA to 3 A
AC Current Range100 µA to 3 A
Resistance Range100 MΩ
Frequency Range3 Hz to 300 kHz
Frequency Accuracy0.01% + 1 count
Capacitance Range1 nF to 100 µF
Capacitance Accuracy1% + 5 counts
ConnectivityUSB, GPIB (optional)
Dimensions88 mm
Power Supply100 V to 240 V AC, 50/60 Hz
DC Voltage Accuracy±0.012% + 2 counts

Related product manuals