Series 3700A System Switch/Multimeter Reference Manual Section 11: TSP command reference
3700AS-901-01 Rev. D/June 2018 11-21
Example
reset()
testData = dmm.makebuffer(1000)
testData.collectchannels = 1
dmm.nplc = 0.5
dmm.range = 0
dmm.configure.set("Dcv_100mV")
dmm.setconfig("slot2", "Dcv_100mV")
scan.create("2035:2040")
scan.execute(testData)
print(testData.channels[1])
printbuffer(1, 6, testData.channels)
This example creates a reading buffer named
testData, configures the buffer to collect
channel data, sets and saves the DMM
configuration, creates a scan list, and then
runs the scan.
The print() command then outputs the first
measurement channel.
Output:
2035+
The printbuffer() command then outputs
the channels for measurements 1 to 6 in the
reading buffer.
Output:
2035+, 2036+, 2037+, 2038+, 2039+,
2040+
Also see
bufferVar.clear() (on page 11-21)
bufferVar.collectchannels (on page 11-23)
Reading buffers (on page 6-7, on page 6-1)
bufferVar.clear()
This function empties the buffer.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
bufferVar.clear()
bufferVar
The reading buffer
Details
This function clears all readings and statistics from the specified buffer (for example, time, status,
channels, and units) from the specified buffer.

Table of Contents

Other manuals for Keithley 3721-ST

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 3721-ST and is the answer not in the manual?

Keithley 3721-ST Specifications

General IconGeneral
TypeMultimeter
Model3721-ST
ManufacturerKeithley
Measurement FunctionsDC Voltage, AC Voltage, DC Current, AC Current, Resistance, Frequency, Temperature
Relay TypeElectromechanical
Temperature Range-200°C to 1820°C
AccuracyVaries by function and range
ResolutionVaries by function and range
DisplayLCD
InterfaceGPIB, RS-232
TopologyMultiplexer
Maximum Switching Voltage300 V
Maximum Switching Current1 A
Input Impedance>10 GΩ (DC Voltage ranges)