Series 3700A System Switch/Multimeter Reference Manual Section 11: TSP command reference
3700AS-901-01 Rev. D/June 2018 11-421
tsplink.master
This attribute reads the node number assigned to the master node.
Usage
masterNodeNumber = tsplink.master
The node number of the master node
Details
After doing a TSP-Link reset (tsplink.reset()), use this attribute to access the node number of
the master in a set of instruments connected over TSP-Link.
Example
LinkMaster = tsplink.master
Store the TSP-Link master
node number in a variable
called LinkMaster.
Also see
tsplink.reset() (on page 11-423)
tsplink.node
This attribute defines the node number.
Usage
nodeNumber = tsplink.node
tsplink.node = nodeNumber
The node number of the instrument or enclosure (1 to 64)
Details
This command sets the TSP-Link node number and saves the value in nonvolatile memory.
Changes to the node number do not take effect until tsplink.reset() from an earlier TSP-Link
instrument is executed on any node in the system.
Each node connected to the TSP-Link system must be assigned a different node number.
Example
Sets the TSP-Link node for this instrument to number 3.
Also see
tsplink.reset() (on page 11-423)
tsplink.state (on page 11-424)