Section 11: TSP command reference Series 3700A System Switch/Multimeter Reference Manual
11-322 3700AS-901-01 Rev. D/June 2018
scan.nobufferbackground()
This function starts a scan in background mode and specifies that no reading buffer is used during scanning.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
state, scancount, stepcount = scan.nobufferbackground()
state
The result of scanning:
â–ª scan.EMPTY or 0
â–ª scan.BUILDING or 1
â–ª scan.RUNNING or 2
â–ª scan.ABORTED or 3
â–ª scan.FAILED or 4
â–ª scan.FAILED_INIT or 5
â–ª scan.SUCCESS or 6
scancount
The present number of scans completed
stepcount
The present number of steps completed
Details
Before using this command, use scan.create(), scan.add(), and scan.addimagestep() to
set up scan elements. If a reading buffer is specified, an error is generated.
To view the scan status, use scan.state().
To run a scan in the background with a reading buffer, see scan.background() (on page 11-313).
Example
scan.nobufferbackground()
Run the scan in the background with no
reading buffer.
Also see
scan.add() (on page 11-309)
scan.background() (on page 11-313)
scan.create() (on page 11-315)
scan.execute() (on page 11-317)
scan.list() (on page 11-318)
scan.nobufferexecute() (on page 11-323)
scan.state() (on page 11-326)
Scanning and triggering (on page 5-23)

Table of Contents

Other manuals for Keithley 3721-ST

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 3721-ST and is the answer not in the manual?

Keithley 3721-ST Specifications

General IconGeneral
BrandKeithley
Model3721-ST
CategoryMultimeter
LanguageEnglish

Related product manuals