IEEE-488 Reference
3-146
ger Link output line. For all other control source selections, the output trigger pulse is available
at the METER COMPLETE Output connector. The output trigger in these two layers is disabled
when the source bypass is disabled.
The output trigger in the measure layer is always enabled and occurs after the device action.
With ACCeptor selected, the bypass is disabled, This simply means that operation will not pro-
ceed through the specified layer until the appropriate event occurs.
:ASYNchronous commands
:ILINe <NRf>
:ARM[:SEQuence[1]][:LAYer[1]]:TCONfigure:ASYNchronous:ILINe <NRf> Select input trigger line; arm
layer
:ARM[:SEQuence[1]]:LAYer2:TCONfigure:ASYNchronous:ILINe <NRf> Select input trigger line; scan
layer
:TRIGger[:SEQuence[1]]:TCONfigure:ASYNchronous:ILINe <NRf> Select input trigger line; mea-
sure layer
Parameters <NRf> = 1 Line #1 <NRf> = 4 Line #4
2 Line #2 5 Line #5
3 Line #3 6 Line #6
Query :ILINe? Query programmed input line #
Description These commands are used to select an input line for the asynchronous Trigger Link of the spec-
ified layer. Keep in mind that asynchronous Trigger Link input and output (see :OLINe) cannot
share the same line. If you assign the input to a line that is already being used for the output, an
error occurs and the command is ignored.
:OLINe <NRf>
:ARM[:SEQuence[1]][:LAYer[1]]:TCONfigure:ASYNchronous:OLINe <NRf> Select output trigger line;
arm layer
:ARM[:SEQuence[1]]:LAYer2:TCONfigure:ASYNchronous:OLINe <NRf> Select output trigger line;
scan layer
:TRIGger[:SEQuence[1]]:TCONfigure:ASYNchronous:OLINe <NRf> Select output trigger line;
measure layer
Parameters <NRf> = 1 Line #1 <NRf> = 4 Line #4
2 Line #2 5 Line #5
3 Line #3 6 Line #6
Query :OLINe? Query programmed output line #
Description These commands are used to select an output line for the asynchronous Trigger Link of the spec-
ified layer. Keep in mind that asynchronous Trigger Link input and output cannot share the same
line. If you assign the output to a line that is already being used for the input, an error occurs
and the command is ignored.

Table of Contents

Other manuals for Keithley 6517A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6517A and is the answer not in the manual?

Keithley 6517A Specifications

General IconGeneral
BrandKeithley
Model6517A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals