IEEE-488 Reference
3-151
3.25.6 :SVOLtage <NRf>:TSEQuence:CLEakage:SVOLtage <NRf> Bias voltage; cap leakage test
:TSEQuence:CIResistance:SVOLtage <NRf> Bias voltage; cable insulation res test
:TSEQuence:RVCoefficient:SVOLtage[1] <NRf> Bias voltage 1; Res voltage coef test
:TSEQuence:RVCoefficient:SVOLtage2 <NRf> Bias voltage 2; Res voltage coef test
:TSEQuence:SRESistivity:SVOLtage <NRf> Bias voltage; Sur resistivity test
:TSEQuence:VRESistivity:SVOLtage <NRf> Bias voltage; Vol resistivity test
:TSEQuence:SIResistance:SVOLtage <NRf> Bias voltage; Sur insulation res test
Parameters <NRf> = -100.0 to 100.0 Volts; 100V range
-1000 to 1000 Volts; 1000V range
Query :SVOLtage? Query soak voltage
Description This configuration command is used for the following tests:
Capacitor Leakage Test
Cable Insulation Resistance Test
Resistor Voltage Coefficient Test
Surface Resistivity Test
Volume Resistivity Test
Surface Insulation Resistance Test
This command is used to specify the bias voltage for the specified test. Note that the Resistor
Voltage Coefficient Test uses two bias voltages; bias voltage 1 and bias voltage 2.
3.25.7 :STIMe <NRf> :TSEQuence:SRESistivity:STIMe <NRf> Bias time; surface resistivity test
:TSEQuence:VRESistivity:STIMe <NRf> Bias time; volume resistivity test
:TSEQuence:SIResistance:STIMe <NRf> Bias time; sur insul resistance test
:TSEQuence:STSWeep:STIMe <NRf> Bias time; staircase sweep test
Parameters :STIMe <NRf> 0 to 9999.9(Staircase Sweep test) Delay in seconds
0 to 99999.9(all others)Delay in seconds
Query :STIMe? Query soak time
Description This configuration command is used for the following tests:
Surface Resistivity Test
Volume Resistivity Test
Surface Insulation Resistance Test
Staircase Sweep Test
This command is used to specify the bias time for the specified test.

Table of Contents

Other manuals for Keithley 6517A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6517A and is the answer not in the manual?

Keithley 6517A Specifications

General IconGeneral
BrandKeithley
Model6517A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals