IEEE-488 Reference
3-63
:MA2Factor <NRf>
:CALCulate[1]:KMATh:MA2Factor <NRf> Specify "a2" factor
Parameters <NRf> = -9.999999e20 to +9.999999e20
Query :MA2Factor? Query "a2" factor
Description This command is used to specify the "a2" factor for the polynomial calculation.
:PERCent <NRf>
:CALCulate[1]:KMATh:PERCent <NRf> Specify Percent target value
Parameters <NRf> = -9.999999e35 to +9.999999e35 Specify target value for Percent calculation.
Query :PERCent? Query Percent target value
Description This command is used to specify the target value for the Percent calculation.
:REFerence <NRf>
:CALCulate[1]:REFerence <NRf> Set reference value for ratio, deviation and percent deviation
Parameters <NRf> = -9.999999e35 to +9.999999e35 Specify reference value
Query :REFerence? Query reference value
Description This command is used to specify the reference value for the ratio, deviation and percent devia-
tion calculation.
:STATe <b>
:CALCulate[1]:STATe <b> Control CALC 1
Parameters <b> = 0 or OFF Disable CALC 1 calculation
1 or ONEnable CALC 1 calculation
Query :STATe? Query state (on or off) of CALC 1
Description This command is used to enable or disable the CALC 1 calculation. When enabled, each instru-
ment reading will reflect the selected calculation (see :FORMat).
:DATA commands
[:LATest]?
:CALCulate[1]:DATA[:LATest]? Read CALC 1 result
Description This query command is used to read the result of the CALC 1 calculation. If CALC 1 is disabled
or NONE is selected, the "raw" reading will instead be read.

Table of Contents

Other manuals for Keithley 6517A

Related product manuals