IEEE-488 Reference
3-153
3.25.11 :MTIMe <NRf> :TSEQuence:SRESistivity:MTIMe <NRf> Measure time; surface resistivity test
:TSEQuence:VRESistivity:MTIMe <NRf> Measure time; volume resistivity test
:TSEQuence:SIResistance:MTIMe <NRf> Measure time; surface insulation resistance test
:TSEQuence:ALTPolarity:MTIMe <NRf> Measure time; alternating polarity test
Parameters <NRf> = 0 to 9999.9 (seconds)
Note: <NRf> = 0.5 to 9999.9 (seconds) for Alternating Polarity Resistance/Resistivity Test.
Query :MTIMe? Query measure time
Description This configuration command is used for the following tests:
Surface Resistivity Test
Volume Resistivity Test
Surface Insulation Resistance Test
Alternating Polarity Resistance/Resistivity Test
This command is used to specify the measure time for the specified test.
3.25.12 :HLEVel <NRf> :TSEQuence:SQSWeep:HLEVel <NRf> High level voltage; square wave sweep test
Parameters <NRf> = -100.0 to 100.0 Volts; 100V range
-1000 to 1000 Volts; 1000V range
Query :HLEVel? Query high level voltage
Description This configuration command is used for the following test:
Square Wave Sweep Test
This command is used to specify the high level voltage for the square wave sweep test.
3.25.13 :HTIMe <NRf> :TSEQuence:SQSWeep:HTIMe <NRf> High level time; square wave sweep test
Parameters <NRf> = 0 to 9999.9 (seconds)
Query :HTIMe? Query high level time
Description This configuration command is used for the following tests:
Square Wave Sweep Test
This command is used to specify the high level time for the square wave sweep test.
3.25.14 :LLEVel <NRf> :TSEQuence:SQSWeep:LLEVel <NRf> Low level voltage; square wave sweep test
Parameters <NRf> = -100.0 to 100.0 Volts; 100V range
-1000 to 1000 Volts; 1000V range
Query :LLEVel? Query low level voltage
Description This configuration command is used for the following test:
Square Wave Sweep Test
This command is used to specify the low level voltage for the square wave sweep test.