Page #304 background imageLoading...
Page #304 background image
IEEE-488 Reference
3-153
3.25.11 :MTIMe <NRf> :TSEQuence:SRESistivity:MTIMe <NRf> Measure time; surface resistivity test
:TSEQuence:VRESistivity:MTIMe <NRf> Measure time; volume resistivity test
:TSEQuence:SIResistance:MTIMe <NRf> Measure time; surface insulation resistance test
:TSEQuence:ALTPolarity:MTIMe <NRf> Measure time; alternating polarity test
Parameters <NRf> = 0 to 9999.9 (seconds)
Note: <NRf> = 0.5 to 9999.9 (seconds) for Alternating Polarity Resistance/Resistivity Test.
Query :MTIMe? Query measure time
Description This configuration command is used for the following tests:
Surface Resistivity Test
Volume Resistivity Test
Surface Insulation Resistance Test
Alternating Polarity Resistance/Resistivity Test
This command is used to specify the measure time for the specified test.
3.25.12 :HLEVel <NRf> :TSEQuence:SQSWeep:HLEVel <NRf> High level voltage; square wave sweep test
Parameters <NRf> = -100.0 to 100.0 Volts; 100V range
-1000 to 1000 Volts; 1000V range
Query :HLEVel? Query high level voltage
Description This configuration command is used for the following test:
Square Wave Sweep Test
This command is used to specify the high level voltage for the square wave sweep test.
3.25.13 :HTIMe <NRf> :TSEQuence:SQSWeep:HTIMe <NRf> High level time; square wave sweep test
Parameters <NRf> = 0 to 9999.9 (seconds)
Query :HTIMe? Query high level time
Description This configuration command is used for the following tests:
Square Wave Sweep Test
This command is used to specify the high level time for the square wave sweep test.
3.25.14 :LLEVel <NRf> :TSEQuence:SQSWeep:LLEVel <NRf> Low level voltage; square wave sweep test
Parameters <NRf> = -100.0 to 100.0 Volts; 100V range
-1000 to 1000 Volts; 1000V range
Query :LLEVel? Query low level voltage
Description This configuration command is used for the following test:
Square Wave Sweep Test
This command is used to specify the low level voltage for the square wave sweep test.

Table of Contents

Other manuals for Keithley 6517A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6517A and is the answer not in the manual?

Keithley 6517A Specifications

General IconGeneral
BrandKeithley
Model6517A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals